INTUITIVE SURGICAL, INC ENDOWRIST;DAVINCI SI; MEGASUTURECUT ND Back to Search Results
Model Number 420309-05
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) received the instrument involved with this complaint and completed the device evaluation.Failure analysis could not confirm the customer reported failure.The instrument was placed on an in-house system and the instrument passed recognition and engagement tests.However, the instrument was found to have a broken pitch cable at the clevis hub.The broken cable segment that contains the crimp was still installed in the clevis.Isi has conducted a device history record (dhr) review for this device and did not find any non-conformances that were related to this reported event.The customer reported complaint does not itself constitute an mdr reportable event; however, the broken pitch cable found during failure analysis could cause or contribute to an adverse event if the malfunction were to recur.
 
Event or Problem Description
It was reported that prior to the start of a da vinci-assisted total benign hysterectomy procedure, the mega suture cut instrument could not be recognized by the system.There was no report of fragment(s) falling inside the patient, patient harm, adverse outcome or injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST;DAVINCI SI
Common Device Name
MEGASUTURECUT ND
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
950 kifer rd.
sunnyvale
MDR Report Key7574997
Report Number2955842-2018-10263
Device Sequence Number274944
Product Code NAY
UDI-Device Identifier00886874111796
UDI-Public(01)00886874111796(10)N10171115
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K050369
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number420309-05
Device Lot NumberN10171115 630
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/15/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/21/2018
Initial Report FDA Received Date06/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-