| Model Number |
420309-05 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the instrument involved with this complaint and completed the device evaluation.Failure analysis could not confirm the customer reported failure.The instrument was placed on an in-house system and the instrument passed recognition and engagement tests.However, the instrument was found to have a broken pitch cable at the clevis hub.The broken cable segment that contains the crimp was still installed in the clevis.Isi has conducted a device history record (dhr) review for this device and did not find any non-conformances that were related to this reported event.The customer reported complaint does not itself constitute an mdr reportable event; however, the broken pitch cable found during failure analysis could cause or contribute to an adverse event if the malfunction were to recur.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to the start of a da vinci-assisted total benign hysterectomy procedure, the mega suture cut instrument could not be recognized by the system.There was no report of fragment(s) falling inside the patient, patient harm, adverse outcome or injury.
|
| |
|