• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. VERSADIAL HEAD IMPACTOR; PROSTHESIS, SHOULDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. VERSADIAL HEAD IMPACTOR; PROSTHESIS, SHOULDER Back to Search Results
Model Number N/A
Device Problem Degraded (1153)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/25/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Foreign: (b)(6).Complaint sample was evaluated and the reported event was confirmed.The impactor was returned for evaluation.As returned, the threads on the strike plate and the impactor tip fractured.Dhr was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that at the beginning of a right total shoulder arthroplasty, the distal part of the impactor was stripped.Another impactor was used to complete the surgery.There was no injury to the patient.The impactor did not come in contact with the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSADIAL HEAD IMPACTOR
Type of Device
PROSTHESIS, SHOULDER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7575013
MDR Text Key110359968
Report Number0001825034-2018-03818
Device Sequence Number1
Product Code HWA
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 06/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number407297
Device Lot Number046230
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/05/2018
Initial Date FDA Received06/06/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/30/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-