| Model Number |
EM210RF |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation was not performed, could not perform the temperature test as the motor in the returned drill would not run.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Cusrtomer reported the em210 heated up.No patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected information.The actual/suspect hand piece has tested for heat.The ambient starting temperature was 82.4 degrees f and rose to 88.7 degrees f after running for one minute.The maximum allowable temperature after running for one minute is 118 degrees f.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Customer reported the drill heated up.There was no patient injury.Analysis did not confirm a heating issue.
|
| |
|