| Catalog Number |
9734913 |
| Medical Device Problem Code |
Bent (1059)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
05/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No patient information provided as no patient was involved in this concern.Lot no.And manufacture date not available for this instrument at time of filing.No parts have been returned to the manufacturer for evaluation.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that, while outside of a procedure, the thoracic probe for the site was bent.There was no patient present when this issue was identified.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: device lot number provided.The probe was returned to the manufacturer for evaluation.Testing found that the tip of the probe was not bent.However, it was noted that impact marks at the back end led to fit issues with trackers.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: unique device identification (udi) and device manufacture date provided.
|
| |
|