• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) INST KIT PROBE THORACIC ASSY; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) INST KIT PROBE THORACIC ASSY; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9734913
Medical Device Problem Code Bent (1059)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 05/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Lot no.And manufacture date not available for this instrument at time of filing.No parts have been returned to the manufacturer for evaluation.
 
Event or Problem Description
Medtronic received information that, while outside of a procedure, the thoracic probe for the site was bent.There was no patient present when this issue was identified.No additional information was provided.
 
Additional Manufacturer Narrative
Additional information: device lot number provided.The probe was returned to the manufacturer for evaluation.Testing found that the tip of the probe was not bent.However, it was noted that impact marks at the back end led to fit issues with trackers.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Additional Manufacturer Narrative
Additional information: unique device identification (udi) and device manufacture date provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INST KIT PROBE THORACIC ASSY
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
MDR Report Key7575254
Report Number1723170-2018-02550
Device Sequence Number13989260
Product Code OLO
UDI-Device Identifier00643169636095
UDI-Public00643169636095
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K124004
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Catalogue Number9734913
Device Lot Number170606
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2018
Initial Date Received by Manufacturer 05/14/2018
Supplement Date Received by Manufacturer08/07/2018
08/28/2018
Initial Report FDA Received Date06/06/2018
Supplement Report FDA Received Date08/28/2018
09/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-