• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN 840 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN 840 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 840
Medical Device Problem Codes Thermal Decomposition of Device (1071); Smoking (1585)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic has not received the suspect device/component from the customer for evaluation nor has the field service engineer evaluated the device.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the 840 ventilator failed the short self test (sst).The ventilator experienced smoke and a burning smell.The ventilator was not in use on a patient at the time of the reported event.
 
Additional Manufacturer Narrative
Device evaluation summary: the field service engineer evaluated the ventilator and replaced the touch frame cable.The ventilator passed all testing per manufacturer's specification and was returned to the customer for clinical use.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
840 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
MDR Report Key7575356
Report Number8020893-2018-00251
Device Sequence Number11206546
Product Code CBK
UDI-Device Identifier10884521077058
UDI-Public10884521077058
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K151252
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number840
Device Catalogue Number4-840220DIEC-SP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2018
Initial Date Received by Manufacturer 05/15/2018
Supplement Date Received by Manufacturer07/30/2018
08/08/2018
Initial Report FDA Received Date06/06/2018
Supplement Report FDA Received Date08/08/2018
10/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-