Model Number 70-0024-020
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There was no patient involvement.Problem was discovered prior to use on a patient.Model number is xeridiem's part number for the device.Catalog number is part number for xeridiem's exclusive distributor for the device, boston scientific.Codes chosen reflect results and conclusion from evaluation of the returned device and package sealing process for the device.Insufficient clarity of visual work instruction for sealing process identified as likely root cause.Xeridiem is correcting the associated visual work instruction.
 
Event or Problem Description
When the product was being unpacked, one of the 20fr replacement g-tube, straight, was open and the device was unsterile.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOVIVE STANDARD BALLOON REPLACEMENT KIT, STRAIGHT BOLSTER
Common Device Name
GASTROSTOMY TUBE
Manufacturer (Section D)
XERIDIEM MEDICAL DEVICES
4700 s. overland drive
tucson AZ 85714 3430
MDR Report Key7575399
Report Number2025851-2018-00014
Device Sequence Number2222133
Product Code KNT
UDI-Device Identifier00813939020281
UDI-Public(01)00813939020281(17)200315(10)0976103
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K083684
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/15/2020
Device Model Number70-0024-020
Device Catalogue NumberM00582080
Device Lot Number976103
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/30/2018
Initial Report FDA Received Date06/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-