Catalog Number 326666
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/15/2018
Event Type  malfunction  
Manufacturer Narrative
Medical device expiration date: unknown.  device manufacture date: unknown. Investigation results: summary: investigation summary: customer returned photos of a 1/2cc syringe. Customer states that the stopper wasn¿t installed. The photos were examined and exhibited the syringe with the stopper correctly in place in the barrel. Unable to perform dhr check due to unknown lot number. Complaints received for this device and reported condition will continue to be tracked and trended. Information will be captured on trend reports and monitored monthly. Our business team regularly reviews the collected data for identification of emerging trends. Based on the above, no additional investigation and no capa is required at this time conclusion: based on the samples / photo(s) received the investigation concluded: unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure. Root cause: root cause cannot be determined at this time as the issue is unconfirmed.
 
Event Description
It was reported that the stopper wasn't installed on a bd ultra-fine¿ insulin syringe. There was no report of exposure to mucous membranes, serious injury, or medical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameBD ULTRA-FINE¿ INSULIN SYRINGE
Type of DeviceINSULIN SYRINGE WITH NEEDLE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer (Section G)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652845
MDR Report Key7575511
MDR Text Key110473573
Report Number1920898-2018-00408
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K941657
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation
Type of Report Initial
Report Date 06/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/06/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number326666
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/16/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 06/06/2018 Patient Sequence Number: 1
-
-