Model Number 3503190
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 05/15/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
At the time of this report, mentor has received no information regarding explantation or an expected explantation date.It is unknown at this time if the device will be made available for return.As a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A device history record review is in progress.Once completed, a supplemental report will be submitted.Concomitant products: mentor smooth round high profile 190cc saline prosthesis, catalog #3503190, serial #(b)(4).
 
Event or Problem Description
It was reported that a (b)(6) -year-old caucasian female underwent primary breast augmentation with a mentor smooth round high profile 190cc saline prosthesis.On (b)(6) 2018 left sided deflation was observed.At the time of this report, mentor has received no information regarding explantation or an expected explantation date.
 
Additional Manufacturer Narrative
The device history record (dhr) of lot number 5793793 was reviewed on 6/22/2018 and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Additional information was received on april 1, 2020.An explant is scheduled for (b)(6) 2020.2.Is other serious-uncheck 2.Is required intervention-check since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Additional information was received on july 6, 2020.The explant date has rescheduled to (b)(6) 2020.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
The impacted product was received by the failure analysis lab on september 3, 3030.The investigation of the returned device was completed by the failure analysis lab on september 23, 2020.Device evaluation summary: during the visual evaluation, an anomaly was observed on the anterior extending to the posterior view consistent with a crease/fold.A tear on the anterior aspect measuring approximately 0.2 cm was also observed within the crease/fold.The evaluation determined that the loss of shell integrity is consistent with a crease fold deflation of the implant shell, which is a known inherent risk of saline-filled mammary prosthesis.Deflation can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Creating wrinkles or folds should be avoided during implantation or other procedures as it can result in deflation at a later time.Breast implants may also simply wear out over time.Each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Additional information was received on (b)(6) 2020.In addition to the left sided deflation reported initially, right sided capsular contracture was also noted.When the devices were explanted, bilateral prosthesis replacement with 500cc mentor memorygel breast implants was performed.This report relates to the left prosthesis.See 1645337-2020-13284, for contralateral prosthesis report.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR SMOOTH ROUND HIGH PROFILE
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
MDR Report Key7575955
Report Number1645337-2018-03525
Device Sequence Number2222121
Product Code FWM
UDI-Device Identifier00081317001577
UDI-Public00081317001577
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P990075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2020
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 05/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2011
Device Model Number3503190
Device Catalogue Number3503190
Device Lot Number5793793
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/03/2020
Initial Date Received by Manufacturer 05/18/2018
Supplement Date Received by Manufacturer05/18/2018
04/01/2020
07/06/2020
09/03/2020
10/12/2020
Initial Report FDA Received Date06/06/2018
Supplement Report FDA Received Date07/10/2018
04/03/2020
07/24/2020
09/23/2020
11/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age35 YR
-
-