Model Number DF31IT36K
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported the catheter has holes in the hub.The catheter was removed and replaced with a new device.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: dhr was reviewed and no deviations/issues were identified associated with this problem in regards to product materials or during packaging, manufacturing or qc inspection processes.All necessary inspections were performed throughout all the manufacturing, packaging and inspection activities and for raw material utilized in the manufacture of this lot, in regards the problem described.Investigation summary: the sample returned for evaluation was one 31 cm decathlon catheter.Leak testing, examination of the damage, and replication of the damage using non-complainant samples was performed.As the damage is consistent with a guidewire's presence during clamping, the complaint is confirmed as use related.Labeling review:a review of the product instructions for use (ifu) procedural instructions, indications, warnings, precautions, cautions, possible complications and contraindications showed that the product labeling is adequate.The ifu states, ¿4.Flush each lumen with heparin solution prior to insertion and clamp the extension legs.If using stylet, do not clamp the arterial (red) lumen until the arterial insertion stylet and guidewire are removed.Clamping will kink the stylet and prevent guidewire passage.¿.
 
Event or Problem Description
It was reported the catheter has holes in the hub.The catheter was removed and replaced with a new device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DECATHLON DF LONG-TERM HEMODIALYSIS CATHETER INTERMEDIATE TRAY (16F) (DUAL-LUMEN
Common Device Name
CATHETER, HEMODIALYSIS, IMPLANTED
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7576069
Report Number3006260740-2018-01260
Device Sequence Number1698676
Product Code MSD
UDI-Device Identifier00801741013997
UDI-Public(01)00801741013997
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K033294
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDF31IT36K
Device Catalogue NumberDF31IT36K
Device Lot NumberRECN0533
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2018
Initial Date Received by Manufacturer 05/23/2018
Supplement Date Received by Manufacturer08/28/2018
Initial Report FDA Received Date06/06/2018
Supplement Report FDA Received Date09/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age44 YR
Patient Weight100
-
-