Film evaluation summary; the exact cause of the left/contra limb migration out of the left common iliac artery could not be determined from the pre-implant films provided.Films during implant and current ct¿s were not available for review, and the reported migration event could not be assessed.Review of a pre-implant film report revealed that the patient had a 63mm max diameter infrarenal aaa.The proximal neck diameter measured 24mm along a 20mm length, and the neck was moderately angulated to ~55deg and mildly calcified.Both common iliac arteries were moderately tortuous, calcified and short.The left common iliac artery diameter ranged from 21 ¿ 14mm as it coursed distally along a 21mm length.It appears likely that disease progression, with iliac artery dilatation currently to 22 ¿ 17mm in diameter, combined with the short iliac artery length, led to the reported the lack of distal seal and limb migration.If information is provided in the future, a supplemental report will be issued.
|