COVIDIEN EVERFLEX SELF-EXPANDING PERIPHERAL STENT WITH ENTRUST DELIVERY SYSTEM STANDARD; STENT, SUPERFICIAL FEMORAL ARTERY Back to Search Results
Catalog Number EVD35-08-040-080
Medical Device Problem Codes Inaccurate Delivery (2339); Defective Device (2588)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the physician used an everflex entrust device with a 6f sheath and non-mdt.035 guidewire for treatment of a fistula (concentric with no calcium) in a venous axillary vessel.Ifu followed and device prepped without issue.No embolic protection was used the device did not pass through a previously deployed stent.No resistance was encountered when advancing the device no excessive force was used.When the physician attempted to deploy the stent, it jumped forward and was reported to have landed midway through the lesion.The physician stated that wires were sticking out of the area on the entrust delivery handle when he removed the red safety pin.He did not notice these wires until after the stent had been deployed and looked at the delivery system.The physician opted not to place a second stent as there was no flow disruption.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERFLEX SELF-EXPANDING PERIPHERAL STENT WITH ENTRUST DELIVERY SYSTEM STANDARD
Common Device Name
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
MDR Report Key7577135
Report Number2183870-2018-00318
Device Sequence Number11391724
Product Code NIP
UDI-Device Identifier00821684051276
UDI-Public00821684051276
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P110023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/13/2021
Device Catalogue NumberEVD35-08-040-080
Device Lot NumberA612862
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/09/2018
Initial Report FDA Received Date06/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age53 YR
-
-