| Model Number |
ZLB00 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Age/date of birth: unknown, information not provided.Gender/sex: unknown, information not provided.Date of event: unknown; (b)(4).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zlb00 23.5 diopter lens was implanted on (b)(6) 2018 into the patient's operative eye.It was later explanted on (b)(6) 2018 due to a hyperopic surprise.The replacement lens was the same model, but different diopter of 24.5.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation: yes.Returned to manufacturer on: 05/29/2018.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection with the unaided eye shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|