• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS, INC. ADVIA CENTAUR XP TNI-ULTRA ASSAY; TROPONIN IMMUNOASSAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS, INC. ADVIA CENTAUR XP TNI-ULTRA ASSAY; TROPONIN IMMUNOASSAY Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The tube type is sodium heparinized green top tube with gel.The sample tube was initially spun for 6 minutes at 3410 rpms using a swing out centrifulge (gel is flat).The customer confirmed the reagent readypack was well mixed and there was no evidence of reagent clumping in the pack.The customer alleged the sample was fine and no visible clots or fibrin strands noted.The sample was not respun prior to either repeat testing.The siemens customer service engineer (cse) was sent to the customer site for system inspection.The cse performed the total service call and no issues were identified.The cse inspected the specimen and found no issues.The customer ran quality control without any errors or discrepant results.The cause for the discordant tni-ultra result is unknown.Siemens healthcare diagnostics is investigating.The instruction for use (ifu) under the summary and explanation of the test section states the following: "always analyze ctni results in the context of time elapsed since patient presentation to the hospital.Serial sampling is recommended to detect the temporal rise and fall of troponin levels characteristic of mi.In accord with published recommendations, serial testing of ctni at intervals of 2 to 4 hours for up to 12 to 24 hours is suggested to corroborate a single ctni result.An elevated troponin alone is not sufficient to make the diagnosis of mi.Other markers, such as ck-mb and myoglobin, can be used in conjunction with ctni results in aiding the diagnosis of mi.".
 
Event or Problem Description
A falsely elevated advia centaur xp troponin ultra (tni-ultra) result was obtained for a patient sample.The result was reported for the in-patient and questioned by the physician.The patient sample was repeated without respinning and the result was negative.A corrected report was issued.The lead technician repeated the same sample tube in the morning, without respinning the tube, and the result was negative.The corrected result matched the patient's clinical symptoms.Patient treatment was not prescribed or altered.There was no report of adverse health consequences due to the discordant tni-ultra result.
 
Additional Manufacturer Narrative
Siemens filed the initial mdr 1219913-2018-00164 on june 07, 2018.06/08/2018 additional information: post service follow up with the customer confirmed no further issues noted with the quality controls or patient samples for tni-ultra.The cause for the discordant advia centaur xp tni-ultra result is unknown.Preanalytic variables can affect the quality of the sample.While there is insufficient information to determine the cause of the false results, we cannot rule out preanalytic factors as the causative agent.The instrument is performing within specifications.No further evaluation of the device is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA CENTAUR XP TNI-ULTRA ASSAY
Common Device Name
TROPONIN IMMUNOASSAY
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS, INC.
511 benedict avenue
tarrytown NY 10591 5097
MDR Report Key7577824
Report Number1219913-2018-00164
Device Sequence Number284045
Product Code MMI
UDI-Device Identifier00630414516226
UDI-Public00630414516226
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K053020
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/20/2018
Device Model NumberN/A
Device Catalogue Number10317709
Device Lot Number82258130
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/21/2018
Supplement Date Received by Manufacturer06/08/2018
Initial Report FDA Received Date06/07/2018
Supplement Report FDA Received Date06/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No