| Model Number |
88863259-62 |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/21/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during an open procedure, the instrument had a missing needle.The suture package was opened with only 4 strands present.The package was preserved and a new package was opened to complete the case.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.A visual inspection of the returned product noted that there were only four products inside the packaging instead of five.A review of the device history record indicates this product was released meeting all quality release specifications at the time of manufacture.However, a manufacturing fault was identified during product analysis.The root cause of the observed condition was determined to be a result of a manufacturing activity.The missing needle went unnoticed by assembly personnel.Improvements have been initiated to mitigate this condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during an open procedure, the instrument had a missing needle.The suture package was opened with only 4 strands present.The package was preserved and a new package was opened to complete the case.There was no patient injury.
|
| |
|