| Catalog Number |
387.34 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); No Code Available (3191)
|
| Date of Event |
05/09/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device is an instrument and is not implanted/explanted.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a removal of a pediatric osteotomy plate on (b)(6) 2018.The plate removal was a patient¿s choice after a successful healing.During the procedure, there was a locking screw fused into the plate and when the surgeon attempted to remove the screw, the head stripped.The conical extractor was then used but it broke inside the screw.The surgeon wanted to saw the plate to separate it from the screw; however, there was no saw available in the hospital; thus, the patient¿s incision site was closed after about forty-five (45) minutes of trying to remove the plate.Reportedly this event resulted in a forty (40) minute surgical delay.The patient was reported as stable and in good condition.The plate removal was done on unknown date using a saw in another facility.This report is for the conical extraction screw.The plate and screw will be captured under related complaint (b)(4).This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Patient ethnicity is unknown.Code 3191 used to capture forty (40) minute surgical delay.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|