Results: the velocity was fractured approximately 48.5 cm from the hub and support coil winds were exposed.A plastic sleeve was returned over the velocity catheter shaft.Conclusions: evaluation of the returned velocity revealed that the device was fractured.If the device is forcefully retracted against resistance, damage such as this may occur.The root cause of the reported resistance could not be determined.The ace 68 identified in the complaint was not returned for evaluation.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing a thrombectomy procedure in the right middle cerebral artery (mca) using a velocity delivery microcatheter (velocity).During the procedure, the physician used a velocity to bring up a penumbra system ace 68 reperfusion catheter (ace 68).While removing the velocity, the physician experienced resistance and upon removal, it was noticed that the velocity was stretched.The procedure was completed using the same ace 68.There was no report of an adverse effect to the patient.
|