• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 8DFEN
Medical Device Problem Code Air Leak (1008)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, a whistle was felt/heard coming from the tube.The customer reported that there was no patient harm.
 
Additional Manufacturer Narrative
Evaluation summary: the samples were received for evaluation.An inflation/deflation test was performed on each of the four cannulas using a syringe 15cc of air was applied to the cuffs, and it was observed the cuffs inflated and held the air, the samples were left inflated 24 hours and no leaks were observed.Additionally the cuffs, inflation lines and pvt valves were submerged into a container filled with water and no leaks were observed.No failure mode was observed in the received samples.Information has been added to the database and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez,ci 32590
MX  32590
MDR Report Key7578293
Report Number2936999-2018-00376
Device Sequence Number1698154
Product Code JOH
UDI-Device Identifier10884522006675
UDI-Public10884522006675
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K962173
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,foreign,user facilit
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/24/2021
Device Model Number8DFEN
Device Catalogue Number8DFEN
Device Lot Number16K0765JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2018
Initial Date Received by Manufacturer 05/22/2018
Supplement Date Received by Manufacturer10/15/2018
Initial Report FDA Received Date06/07/2018
Supplement Report FDA Received Date11/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age25 YR
Patient Weight47
-
-