| Model Number |
M00518251 |
| Medical Device Problem Code |
Occlusion Within Device (1423)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/22/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
|
| |
|
Event or Problem Description
|
|
Note: this report pertains to one of two devices used during the same procedure.Refer to manufacturer report# 3005099803-2018-01827 for the first interject¿ sclerotherapy needle and 3005099803-2018-01828 for the second interject¿ sclerotherapy needle.It was reported to boston scientific corporation that an interject¿ needle was used in the stomach during a gastroscopy procedure performed on (b)(6) 2018.According to the complainant, during preparation, at the time of injection the liquid did not leave the needle.A second interject ¿ needle was used, and had the same issue of liquid not leaving the needle.The procedure was completed with a different device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
| |
|
Additional Manufacturer Narrative
|
|
An interject needle was returned for analysis.A visual evaluation of the returned device revealed that there were no visual defects noted, the returned device looks under good conditions without evidence of damages.Functional evaluation with the working length formed into two 2¿ diameter loops found the needle would extend and retract without issue.The handle operation was smooth when actuating the device.A second functional inspection was performed, a spare lab syringe was filled with water, several attempts to inject the water were made but they were unsuccessful, the water did not come out from the needle tip.The device was disassembled and an occlusion was found blocking the inner sheath at the proximal side of the needle.The reported complaint that the needle was noticed to be occluded was confirmed.Based on the information available and the analysis performed the complaint will be documented as "manufacturing execution error", the manufacturing process was not executed as validated/as designed.There is an investigation to address this issue.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications at the time of release for distribution.
|
| |
|
Event or Problem Description
|
|
Note: this report pertains to one of two devices used during the same procedure.Refer to manufacturer report# 3005099803-2018-01827 for the first interject sclerotherapy needle and 3005099803-2018-01828 for the second interject sclerotherapy needle.It was reported to boston scientific corporation that an interject needle was used in the stomach during a gastroscopy procedure performed on (b)(6) 2018.According to the complainant, during preparation, at the time of injection the liquid did not leave the needle.A second interject needle was used, and had the same issue of liquid not leaving the needle.The procedure was completed with a different device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
| |
|
Search Alerts/Recalls
|