• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - SPENCER INTERJECT¿; BIOPSY NEEDLE KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - SPENCER INTERJECT¿; BIOPSY NEEDLE KIT Back to Search Results
Model Number M00518251
Medical Device Problem Code Occlusion Within Device (1423)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
Note: this report pertains to one of two devices used during the same procedure.Refer to manufacturer report# 3005099803-2018-01827 for the first interject¿ sclerotherapy needle and 3005099803-2018-01828 for the second interject¿ sclerotherapy needle.It was reported to boston scientific corporation that an interject¿ needle was used in the stomach during a gastroscopy procedure performed on (b)(6) 2018.According to the complainant, during preparation, at the time of injection the liquid did not leave the needle.A second interject ¿ needle was used, and had the same issue of liquid not leaving the needle.The procedure was completed with a different device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Investigation results: an interject needle was returned for analysis.A visual evaluation of the returned device revealed that there were no visual defects noted, the returned device looks under good conditions without evidence of damages.Functional evaluation with the working length formed into two 2¿ diameter loops found the needle would extend and retract without issue.The handle operation was smooth when actuating the device.A second functional inspection was performed, a spare lab syringe was filled with water, several attempts to inject the water were made but they were unsuccessful, the water did not come out from the needle tip.The device was disassembled and an occlusion was found blocking the inner sheath at the proximal side of the needle.The reported complaint that the needle was noticed to be occluded was confirmed.Based on the information available and the analysis performed the complaint will be documented as "manufacturing execution error", the manufacturing process was not executed as validated/as designed.There is an investigation to address this issue.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications at the time of release for distribution.
 
Event or Problem Description
Note: this report pertains to one of two devices used during the same procedure.Refer to manufacturer report# 3005099803-2018-01827 for the first interject¿ sclerotherapy needle and 3005099803-2018-01828 for the second interject¿ sclerotherapy needle.It was reported to boston scientific corporation that an interject¿ needle was used in the stomach during a gastroscopy procedure performed on (b)(6) 2018.According to the complainant, during preparation, at the time of injection the liquid did not leave the needle.A second interject ¿ needle was used, and had the same issue of liquid not leaving the needle.The procedure was completed with a different device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERJECT¿
Common Device Name
BIOPSY NEEDLE KIT
Manufacturer (Section D)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
MDR Report Key7578758
Report Number3005099803-2018-01828
Device Sequence Number2527160
Product Code FCG
UDI-Device Identifier08714729296478
UDI-Public08714729296478
Combination Product (Y/N)N
PMA/510(K) Number
K012864
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/15/2020
Device Model NumberM00518251
Device Catalogue Number1825
Device Lot Number21014959
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2018
Initial Date Received by Manufacturer 05/23/2018
Supplement Date Received by Manufacturer07/29/2018
Initial Report FDA Received Date06/07/2018
Supplement Report FDA Received Date08/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age49 YR
-
-