• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910  PINN MULTIHOLE W/GRIPTION 54MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910  PINN MULTIHOLE W/GRIPTION 54MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Catalog Number 121730054
Device Problems Loss of or Failure to Bond (1068); Fracture (1260); Metal Shedding Debris (1804)
Patient Problems Inadequate Osseointegration (2646); Not Applicable (3189)
Event Date 05/11/2018
Event Type  Injury  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Der states that the cup originally looked good on x-ray and patient was doing well.The surgeon saw patient in clinic a week ago and noticed cup had loosened and that there appeared to be metal debris in the joint.The cup was grossly loose and when he removed it there was a small section of gription that had worn off.Doi: (b)(6) 2016; dor: (b)(6) 2018; unknown side.
 
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : the returned acetabular cup was examined by a depuy material scientist.No evidence of material or manufacturing anomalies were observed.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MULTIHOLE W/GRIPTION 54MM
Type of Device
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910 
700 orthopaedic drive
warsaw IN 46582 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910 
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
kara ditty-bovard
1210 ward avenue
west chester, IN 19380-0988
6103142063
MDR Report Key7579483
MDR Text Key110404127
Report Number1818910-2018-61608
Device Sequence Number1
Product Code LPH
UDI-Device Identifier10603295009993
UDI-Public10603295009993
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071784
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 05/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number121730054
Device Lot NumberC58381
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/18/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/20/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age78 YR
Patient Weight47
-
-