One unused peripheral intravenous catheter was received for analysis.This was not the same product used in the complaint, however this unused sample was from the same lot number.This sample was tested based on the complaint issue.The take off force, or the initial force that separates the tip of the catheter from the needle, as well as the force to advance, the force required to advance the device tab until locking, were both tested.The resulting values were within specification.Without evaluating the actual sample, it cannot be assessed or confirmed whether a quality related issue has resulted in the user's reported difficulties.The device history record was reviewed and no manufacturing or testing issues were found.Based on the evidence, the complaint was confirmed.
|