• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065751763
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/16/2018
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported that there was no phaco power although the handpiece primed and tuned successfully.The handpiece was exchanged but did not resolve the issue.A backup system was brought in and the cataract extraction with intraocular lens implant (iol) was completed with no harm to the patient.
 
Manufacturer Narrative
The system was examined and the reported event was not replicated.The system was tested and found to meet product specifications.A review of the customer¿s complaint history for the last 24 months did not show any previous complaints of this kind against the system.The system was manufactured on june 14, 2014.Based on qa assessment, the product met specifications at the time of release.The system to meet specifications as related to the reported event.Therefore, the root cause of the reported event cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key7580208
MDR Text Key110720074
Report Number2028159-2018-01178
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K121555
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 10/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065751763
Other Device ID Number380657517633
Was Device Available for Evaluation? Yes
Date Manufacturer Received09/28/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-