MAUDE Adverse Event Report: BECTON, DICKINSON & CO. BD VACUTAINER® LITHIUM HEPARINN (LH) 56 USP UNITS BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. BD VACUTAINER® LITHIUM HEPARINN (LH) 56 USP UNITS BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 366667
Medical Device Problem Code Short Fill (1575)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/04/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation summary: bd had not received samples, but photos were provided by the customer facility for investigation.The photos were evaluated and the customer's indicated failure mode for underfilled tubes with the incident lot was observed; however, it was also observed that the stoppers were not seated properly on the tubes.A review of the manufacturing records was completed for the incident lot and no issues were identified.Investigation conclusion: based on evaluation of the customer photos, the customer¿s indicated failure mode for underfilled tubes with the incident lot was observed.Root cause description: based on the investigation, a root cause could not be determined.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that bd vacutainer® lithium heparin (lh) 56 usp units blood collection tubes weren¿t filling properly.No serious injury or medical intervention reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® LITHIUM HEPARINN (LH) 56 USP UNITS BLOOD COLLECTION TUBES
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO.
150 south 1st avenue
broken bow NE 68822
MDR Report Key7580427
Report Number1917413-2018-02055
Device Sequence Number2194863
Product Code JKA
UDI-Device Identifier50382903666671
UDI-Public50382903666671
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
K901449
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other,user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date11/30/2018
Device Catalogue Number366667
Device Lot Number7180900
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/29/2017
Initial Report FDA Received Date06/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-