|
Investigation summary: bd had not received samples, but photos were provided by the customer facility for investigation.The photos were evaluated and the customer's indicated failure mode for underfilled tubes with the incident lot was observed; however, it was also observed that the stoppers were not seated properly on the tubes.A review of the manufacturing records was completed for the incident lot and no issues were identified.Investigation conclusion: based on evaluation of the customer photos, the customer¿s indicated failure mode for underfilled tubes with the incident lot was observed.Root cause description: based on the investigation, a root cause could not be determined.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
|