No medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
|
Manufacturing review: dhr was reviewed and no deviations/issues were identified associated with this problem in regards to product materials or during packaging, manufacturing or qc inspection processes.Investigation summary: all necessary inspections were performed throughout all the manufacturing, packaging and inspection activities and for raw material utilized in the manufacture of this lot, in regards the problem described.Manufacturing records were reviewed (dhr, mrr¿s, scrap and mfg.Process changes) and there were not found evidence that the failure mode reported in this one clearvue isp port was returned for evaluation.Visual, functional, and microscopic evaluations were performed.The investigation is confirmed for inability to aspirate, and the event was found more specifically to be due to occlusion by manufacturing material.The port septum was removed.Microscopic evaluation revealed a large amount of clear material on the bottom of the port chamber.Material was also observed within the passage from the port stem to the port body.Manufacturing evaluation identified the material within the stem as over-mold material.The complaint has been confirmed to be manufacturing related.Labeling review: as this complaint has been determined to be manufacturing related, the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications has been determined to be not applicable.Labeling would not have prevented this issue.
|