• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) UNITS BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) UNITS BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 367964
Medical Device Problem Code Incorrect Or Inadequate Test Results (2456)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation summary: bd received samples from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode for erroneous results with the incident lot was not observed as all product specifications were met.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Bd does not recommend pst tubes be used for blood gases.Investigation conclusion: based on evaluation of the customer samples, the customer¿s indicated failure mode for erroneous results with the incident lot was not observed as all samples met the required specifications.Root cause description: based on the investigation, a root cause could not be determined.The product was found to be in conformance and meet release specifications.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that bd vacutainer® pst¿ gel and lithium heparin (lh) 126 units blood collection tubes had erroneous results.No serious injury or medical intervention.
 
Additional Manufacturer Narrative
Bd received samples from the customer facility for investigation.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformance's during manufacturing of the product.Bd does not recommend pst tubes be used for blood gases.Investigation conclusion: the customer¿s indicated failure mode for erroneous results with the incident lot was not observed as all samples met the required specifications.Root cause description: based on the investigation, a root cause could not be determined.The product was found to be in conformance and meet release specifications.
 
Event or Problem Description
It was reported that bd vacutainer® pst¿ gel and lithium heparinn (lh) 126 units blood collection tubes had erroneous results.No serious injury or medical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) UNITS BLOOD COLLECTION TUBES
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO.
150 south 1st avenue
broken bow NE 68822
MDR Report Key7580917
Report Number1917413-2018-01580
Device Sequence Number19406172
Product Code JKA
UDI-Device Identifier50382903679640
UDI-Public50382903679640
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K945952
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date04/30/2018
Device Catalogue Number367964
Device Lot Number7100542
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2018
Initial Date Received by Manufacturer 01/12/2018
Supplement Date Received by Manufacturer01/12/2018
Initial Report FDA Received Date06/07/2018
Supplement Report FDA Received Date03/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-