Model Number 6996SQ58
Medical Device Problem Codes Fracture (1260); High impedance (1291)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/15/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: dtbb1d1 icd, implanted: (b)(6) 2014.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient's right ventricular (rv) lead had high/undefined pacing impedance, and high rv/superior vena cava (svc) coil impedance.The lead was capped and replaced.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
As well, the lead was fractured.
 
Additional Manufacturer Narrative
Product analysis of the device memory indicated the impedance of the right ventricular defibrillation coil was beyond the expected upper range.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUBQ
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key7581795
Report Number2649622-2018-09567
Device Sequence Number2354991
Product Code LWS
UDI-Device Identifier00613994568021
UDI-Public00613994568021
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/18/2017
Device Model Number6996SQ58
Device Catalogue Number6996SQ58
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/15/2018
Supplement Date Received by Manufacturer05/15/2018
06/08/2018
Initial Report FDA Received Date06/08/2018
Supplement Report FDA Received Date06/08/2018
06/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5071 LEAD,6996SQ58 LEAD,4968 LEAD,5866 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age72 YR
-
-