| Medical Device Problem Codes |
Device Displays Incorrect Message (2591); Fail-Safe Problem (2936)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the reporter contacted animas and reported that call service alarm [064 and 078] had been emitted by the pump more often than expected by the user.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the issue may result in a long term cessation of insulin delivery if the user is unable to resolve the alarm.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 25-jun-2018 with the following findings: a review of the black box verified a call service 064-0001 motor fault.The call service alarm was unable to be reproduced during investigation.The pump cover was removed to find evidence of moisture corrosion on the motor flex connector.
|
| |
|