PRIDE MOBILITY PRODUCTS PRIDE MOBILITY PRODUCTS; ELECTRIC POSITIONING CHAIR Back to Search Results
Model Number LC358XL
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Injury (2348)
Date of Event 06/01/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has not been made available for evaluation at this time.Should further information or the device become available, a follow-up report will then be issued.
 
Event or Problem Description
Alleges customer slid off the chair when it was in the raised position.The foam layer on top of the seat cushion came loose from the back of the chair (velcro used to hold foam in place let go) causing the cushion to come forward and throw off the depth perception of the seat.
 
Additional Manufacturer Narrative
Service technician readjusted the seat pad and resecured the velcro.The unit is working properly.
 
Event or Problem Description
Alleges customer slid off the chair when it was in the raised position.The foam layer on top of the seat cushion came loose from the back of the chair (velcro used to hold foam in place let go) causing the cushion to come forward and throw off the depth perception of the seat.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIDE MOBILITY PRODUCTS
Common Device Name
ELECTRIC POSITIONING CHAIR
Manufacturer (Section D)
PRIDE MOBILITY PRODUCTS
401 york ave
duryea PA 18642
MDR Report Key7582133
Report Number2530130-2018-00070
Device Sequence Number2062780
Product Code INO
UDI-Device Identifier00606509400299
UDI-Public00606509400299
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K070950
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberLC358XL
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/07/2018
Supplement Date Received by Manufacturer06/07/2018
Initial Report FDA Received Date06/08/2018
Supplement Report FDA Received Date10/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-