• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Codes Device Displays Incorrect Message (2591); Fail-Safe Problem (2936)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event or Problem Description
On (b)(6) 2018, the reporter contacted animas and reported that call service alarm [078] had been emitted by the pump more often than expected by the user.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the issue may result in a long term cessation of insulin delivery if the user is unable to resolve the alarm.
 
Additional Manufacturer Narrative
Device evaluation: the device has been returned and evaluated by product analysis on 10-jul-2018 with the following findings: during the investigation, a review of the black box showed evidence of multiple cs078 alarms.The pump¿s rewind, load, and prime steps were performed without any alarms or errors occurring.A 24-hour duration test was successfully completed with no call service alarms being duplicated.The pump¿s cover was removed, and no evidence of internal moisture or defects to the printed circuit board or internal components was found.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
200 lawrence dr
west chester PA 19380 3428
MDR Report Key7582429
Report Number2531779-2018-10646
Device Sequence Number19458125
Product Code OYC
UDI-Device Identifier10840406100181
UDI-Public10840406100181
Combination Product (Y/N)Y
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Other Device ID Number1-MCBC-3762
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2018
Was the Report Sent to FDA? Yes
Device Age22 MO
Initial Date Received by Manufacturer 06/02/2018
Supplement Date Received by Manufacturer06/02/2018
Initial Report FDA Received Date06/08/2018
Supplement Report FDA Received Date07/31/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age22 YR
-
-