| Model Number |
CAM02 |
| Medical Device Problem Code |
Invalid Sensing (2293)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/16/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Camino monitor displaying half the actual reading.Patient information: (b)(6).Incident date: (b)(6) 2018.Was there any patient injury/ death alleged: no.Any revisions/medical interventions required: monitor could not be connected to epic (emr) / local monitor as data incorrect - values had to be manually charted.Was the patient (s) prepped for surgery: no.Any delay in surgery due to late receipt of product: no.Any delay in surgery due to product problem(s): no.
|
| |
|
Additional Manufacturer Narrative
|
|
On (b)(6) 2018 the customer reported the following: (b)(4) camino monitor displaying half the actual readings.Patient information: (b)(6) year old male; (b)(6) kg.Incident date: (b)(6) 2018.Was there any patient injury/ death alleged: no.Any revisions/medical interventions required: monitor could not be connected to epic.(emr) / local monitor as data incorrect - values had to be manually charted.Was the patient (s) prepped for surgery: no.Any delay in surgery due to late receipt of product: no.Any delay in surgery due to product problem(s): no.Upon additional followup on 21 august 2018 the customer informed us that incorrect readings were due to user error-customer was using the wrong adapter.After customer switched to the correct adaper the issue resolved.Device will not be returned.No further follow-up needed.This complaint is now deemed closed.
|
| |
|
Search Alerts/Recalls
|