• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Model Number H7493926024270
Medical Device Problem Codes Bent (1059); Difficult to Remove (1528)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device is a combination product.(b)(4).Device evaluated by mfr: the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
 
Event or Problem Description
It was reported that removal difficulties and stent damage occurred.A 2.75 x 24mm synergy ii drug-eluting stent was advanced to treat the lesion.The device was removed to perform further dilatation.Upon removal, the device snagged on the guide catheter and the stent started to pull from the balloon and appeared to have folded over on itself.The guide catheter was pulled into the aorta and through wire manipulation, the device was removed.The procedure was completed with another stent of the same size.There were no patient complications.
 
Event or Problem Description
It was reported that removal difficulties and stent damage occurred.A 2.75 x 24mm synergy ii drug-eluting stent was advanced to treat the lesion.The device was removed to perform further dilatation.Upon removal, the device snagged on the guide catheter and the stent started to pull from the balloon and appeared to have folded over on itself.The guide catheter was pulled into the aorta and through wire manipulation, the device was removed.The procedure was completed with another stent of the same size.There were no patient complications.
 
Additional Manufacturer Narrative
Device avail.For eval, returned to mfr.On, device returned to mfr., device evaluated by mfr., eval summary attached, method codes, result codes, conclusion codes updated.Device evaluated by mfr.: synergy ii us mr 2.75 x 24mm stent delivery system was returned for analysis.A visual examination of the crimped stent identified damage.The proximal end of the stent was bunched and pushed towards the mid-section of the stent.The distal section of the stent was undamaged.The undamaged distal crimped stent outer diameter was measured and was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the device revealed multiple hypotube kinks.An examination of the shaft polymer extrusion found no issues.The bi-component bond showed no signs of damage.The tip was visually and microscopically examined and no issues were noted.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is caused by other device as another device/drug/subsequent procedure caused the complaint event.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7583640
Report Number2134265-2018-05044
Device Sequence Number11391596
Product Code NIQ
UDI-Device Identifier08714729840220
UDI-Public08714729840220
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/03/2018
Device Model NumberH7493926024270
Device Catalogue Number39260-2427
Device Lot Number0020861254
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/22/2018
Initial Date Received by Manufacturer 05/16/2018
Supplement Date Received by Manufacturer07/11/2018
Initial Report FDA Received Date06/08/2018
Supplement Report FDA Received Date07/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-