• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL (BD WEST) MEDICAL SURGICAL BD ¿ PRE-FILLED NORMAL SALINE SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL (BD WEST) MEDICAL SURGICAL BD ¿ PRE-FILLED NORMAL SALINE SYRINGE Back to Search Results
Catalog Number 306559
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/20/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Fda notified: the initial reporter also notified the fda (b)(4).Investigation summary: as no samples or pictures were received for evaluation, our quality team was unable to confirm the reported defect of pump incompatibility.As no lot number was provided, a device history record review could not be completed.It is important to note that the bd franklin flush syringe (product code 306500) does not claim pump compatibility.Bd does not recommend use of this product with a pump.The current controls for syringes including incoming, in-process, and final inspections will remain in place.Investigation conclusion: complaint: pump occlusion and plunger movement difficult.No sample received.A dhr could not be performed as no lot # was provided.There have been zero prior compalints for plunger movement difficult in the past year.There have been similar complaints for "pump alarm/occlusion/incompatibility" within the past year.Note: the bd franklin flush syringe does not claim pump compatibility.Bd does not recommend use of this product with a pump.Bd was not able to duplicate or confirm the customer's indicated failure mode.The customer's indicated failure mode of pump alarm/occlusion and plunger movement could not be confirmed as no sample or photo was provided.Based on the investigation, root cause was not found.No formal corrective action is required at this time.The current controls for syringes including incoming, in-process and final inspections will remain in place.Root cause description based on the investigation, root cause was not found.
 
Event or Problem Description
It was reported that during use a bd ¿ pre-filled normal saline syringe malfunctioned as " the medfusion syringe pumps are having trouble recognizing the loml bd pre- filled flushes.The pump will read "invalid syringe size" and reseating the flush does not work.The change in the plunger when posiflush came out caused increased drag and thereby increased pressure upon deployment of the fluid." there was no report of exposure, injury or medical intervention needed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ¿ PRE-FILLED NORMAL SALINE SYRINGE
Common Device Name
SYRINGE
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
MDR Report Key7583730
Report Number1911916-2018-00318
Device Sequence Number18876590
Product Code FMF
UDI-Device Identifier30382903065593
UDI-Public30382903065593
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K141311
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Risk Manager
Type of Report Initial
Report Date (Section B) 06/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Nurse
Device Catalogue Number306559
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/20/2017
Initial Report FDA Received Date06/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-