Patient information was unavailable from the site.Udi and manufacture date not available for this instrument at time of filing.A medtronic representative went to the site to test the equipment.The system failed the hardware checkout but passed the software and instrument checkout.The axiem box was changed and it resolved the issue.The imaging system then passed the system checkout and was found to be fully functional.The axiem system controller was returned to the manufacturer for analysis.Analysis found that the reported issue was confirmed.The axiem, emitter and tool ports, all showed a red status.The 12vdc power supply shows a ¿fault¿¿.As well as, multiple other drive faults displayed.Analysis found that the reported event was related to an electrical issue.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.The field generator was returned to the manufacturer for analysis.The issue could not be duplicated.There was no fault found with the field generator.
|