| Model Number |
24-473-70-07 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/21/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Reference exemption number e2017029.
|
| |
|
Event or Problem Description
|
|
It was reported that two twist drills broke.All pieces were removed.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).An investigation was performed using a stereo microscope revealed tensile cracks.Further observation determined there were no indications of material or manufacturing defects.The complaint percentage was calculated, and it is determined that the complaint percentage falls within the design risk limits adhered to at klm.During the investigation the product lot number identified was reviewed in the device history records.The dhr review showed no discrepancies or anomalies.The investigation results conclude that the root cause for breakages were due to structural failure of the device due to mechanical overload.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
|
| |
|