• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS POWERPICC SV CATHETER WITH SHERLOCK TIP LOCATION SYSTEM (TLS) STYLET 4F MAXIMAL; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS POWERPICC SV CATHETER WITH SHERLOCK TIP LOCATION SYSTEM (TLS) STYLET 4F MAXIMAL; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number 9274108D
Health Effect - Clinical Code Thrombus (2101)
Date of Event 05/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of recp1588 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported that the picc was placed on (b)(6) 2018 "svc - cavoatrial junction." the picc was "cut at 32 cm, but inserted at 33 cm, plus one - up to the hub." on (b)(6) 2018, the purple lumen became occluded.The nurse successfully used alteplase to dissolve the clot.
 
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of an occluded catheter was inconclusive because the reported event could not be verified in the vad field assurance laboratory.The product returned for evaluation was one 4fr d/l powerpicc sv catheter.Usage residues were observed throughout the sample.The catheter terminated at the 32cm depth marking.An attempt to infuse water through the sample using a 12ml syringe revealed both lumens to be patent to infusion.Blood residue was cleared from both lumens; however, no resistance to infusion was observed.No occlusions were observed; however, the event description indicated that the reported occlusion was successfully cleared using alteplase.A cleared occlusion could not be verified.Consequently this complaint is inconclusive at this time.Blood product occlusion can occur if the catheter is not maintained properly.The product ifu provides guidance for catheter maintenance using heparinized saline.A lot history review (lhr) of recp1588 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported that the picc was placed on (b)(6) 2018 "svc - cavoatrial junction." the picc was "cut at 32 cm, but inserted at 33 cm, plus one - up to the hub." on (b)(6) 2018, the purple lumen became occluded.The nurse successfully used alteplase to dissolve the clot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC SV CATHETER WITH SHERLOCK TIP LOCATION SYSTEM (TLS) STYLET 4F MAXIMAL
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7584502
Report Number3006260740-2018-01314
Device Sequence Number1383913
Product Code LJS
UDI-Device Identifier00801741035678
UDI-Public(01)00801741035678
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
K102159
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 06/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9274108D
Device Catalogue Number9274108D
Device Lot NumberRECP1588
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/07/2018
Event Location Hospital
Initial Date Received by Manufacturer 06/07/2018
Supplement Date Received by Manufacturer06/12/2018
Initial Report FDA Received Date06/08/2018
Supplement Report FDA Received Date06/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age58 YR
Patient Weight53
-
-