BARD ACCESS SYSTEMS POWERPICC SV CATHETER WITH SHERLOCK TIP LOCATION SYSTEM (TLS) STYLET 4F MAXIMAL; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
|
Back to Search Results |
|
| Model Number |
9274108D |
| Health Effect - Clinical Code |
Thrombus (2101)
|
| Date of Event |
05/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of recp1588 showed no other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported that the picc was placed on (b)(6) 2018 "svc - cavoatrial junction." the picc was "cut at 32 cm, but inserted at 33 cm, plus one - up to the hub." on (b)(6) 2018, the purple lumen became occluded.The nurse successfully used alteplase to dissolve the clot.
|
| |
|
Additional Manufacturer Narrative
|
|
The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of an occluded catheter was inconclusive because the reported event could not be verified in the vad field assurance laboratory.The product returned for evaluation was one 4fr d/l powerpicc sv catheter.Usage residues were observed throughout the sample.The catheter terminated at the 32cm depth marking.An attempt to infuse water through the sample using a 12ml syringe revealed both lumens to be patent to infusion.Blood residue was cleared from both lumens; however, no resistance to infusion was observed.No occlusions were observed; however, the event description indicated that the reported occlusion was successfully cleared using alteplase.A cleared occlusion could not be verified.Consequently this complaint is inconclusive at this time.Blood product occlusion can occur if the catheter is not maintained properly.The product ifu provides guidance for catheter maintenance using heparinized saline.A lot history review (lhr) of recp1588 showed no other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported that the picc was placed on (b)(6) 2018 "svc - cavoatrial junction." the picc was "cut at 32 cm, but inserted at 33 cm, plus one - up to the hub." on (b)(6) 2018, the purple lumen became occluded.The nurse successfully used alteplase to dissolve the clot.
|
| |
|