| Model Number |
TV-IL1412120-J |
| Medical Device Problem Codes |
Failure To Adhere Or Bond (1031); Leak/Splash (1354)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Ischemia (1942)
|
| Date of Event |
05/14/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The devices involved in the event will not be returned for evaluation, they remain implanted in the patient.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
Patient initial implanted on (b)(6) 2018 with an ovation to treat an aortic aneurysm.Endologix was made aware of a patient that went into the hospital for reported cold legs (ischemia).Upon review of the ct scan, physician noted the device limbs were placed on the ipsilateral gate and observed a type 3 endoleak.A re-intervention was performed with a non-endologix device on (b)(6) 2018 to resolve the leak.Patient was in good condition post secondary.
|
| |
|