| Model Number |
25199 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned device was evaluated.During evaluation, the device powered on using lab stock batteries; however, an led segment on the led digits failed to fully light up.Internal inspection revealed one instrument board led segment faulty, resulting in display segment being blank.(b)(6).A service history record review reveals that this unit was in the field for over ten (10) months with no previous reported issues prior to this reported event.
|
| |
|
Event or Problem Description
|
|
The customer reported missing led segments.No consequences or impact to patient was reported.
|
| |
|