| Catalog Number |
7640220 |
| Medical Device Problem Code |
Device Slipped (1584)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Neither the device nor applicable imaging films were returned to manufacturer for evaluation therefore we are unable to determine the definitive cause of event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent lumbar spondylolisthesis reduction and internal fixation.Intra-op, the screw slipped.The device came in contact with the patient.No patient complications were reported as a result of the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: macroscopic and visual examination revealed thread crest/flank damage; this damage appears to have initiated near the start of the thread, and is evident around the damaged portion of the thread.Functional evaluation with a sample mas head found the set screw was unable to be fully engaged.The above observations are consistent with misalignment of the mas and set screw threads during construct assembly.If information is provided in the future, a supplemental report will be issued.
|
| |
|