| Catalog Number |
FLR02 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.As a lot/batch was not provided, a device history could not be performed.Additional information was requested and the following was obtained: what part of the device was torn, the iris value seal or the retractor sheath? did any pieces of the device fall into the patient? if yes, how were the pieces retrieved? was there any loss of pneumo due to the tear in the device? how was the procedure completed? was there any patient consequence? response: no further information available.
|
| |
|
Event or Problem Description
|
|
Pulled it lightly, but tear.There were no patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # p9399w.Device analysis: the analysis results of the flr02 found that it was received with the retractor torn.The tear initiated 2 inches below the upper retractor ring and continued toward the lower retractor ring.No conclusion could be reached as to what may have caused the found condition.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
|
| |
|