| Catalog Number |
143140 |
| Medical Device Problem Codes |
Material Integrity Problem (2978); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
Doctor slappey was inserting his first 3.2 x 140mm tibial pin and the tip broke off in the patient¿s bone.He used x-ray to determine the tip was embedded in bone and not floating and decided to leave it in and switch to 4.0 x 140mm pins and completed the case robotically with minimal delay.Doctor slappey admitted the failure could have been due to how he inserted the pin (at an angle and with too much torque) but we still decided to move to 4.0 pins.Tka procedure delayed =15 minutes.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: it was reported that the bone pin broke off during drilling.The doctor decided to leave it in the bone.Product evaluation and results: as per attached x-rays the broken tip was left in the patient.Further evaluation could not performed because the product was unavailable for inspection as the product was not returned.Product history review: review of the product history records indicate 937 devices were manufactured under lot no w52815-2 and accepted into final stock on 09/19/2017.No non-conformances were identified during inspection.Complaint history review: a review of complaints in catsweb and trackwise related to p/n 143140, lot number w52815-2 shows no additional complaints related to the failure in this investigation.Conclusions: the failure mode could not be confirmed because the part was not available for evaluation.If device and/or additional information become available, this investigation will be reopened.Corrective action/preventive action: a review of stryker¿s nc/capa database indicated there have been an ncs (b)(4) and capa (b)(4) are associated with the product and failure mode reported in this event.
|
| |
|
Event or Problem Description
|
|
Doctor (b)(6) was inserting his first 3.2 x 140mm tibial pin and the tip broke off in the patient¿s bone.He used x-ray to determine the tip was embedded in bone and not floating and decided to leave it in and switch to 4.0 x 140mm pins and completed the case robotically with minimal delay.Doctor (b)(6) admitted the failure could have been due to how he inserted the pin (at an angle and with too much torque) but we still decided to move to 4.0 pins.Tka procedure delayed =15 minutes.
|
| |
|