| Catalog Number |
1190300 |
| Medical Device Problem Codes |
Defective Device (2588); Insufficient Device Problem Information (3190); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Codes |
Death (1802); Unspecified Infection (1930); Necrosis (1971); Neuropathy (1983); Pain (1994); Disability (2371)
|
| Date of Event |
05/20/2015
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
Currently, it is unknown to what extent the bard/davol phasix mesh device may have caused or contributed to the events as alleged by the patient¿s attorney.The information provided alleges the patient underwent additional "repair surgery of the hernia and the hernia mesh" with removal of part of the mesh.A sample was not provided for evaluation and no lot number has been provided; therefore a review of the manufacturing records is not possible at this time.No medical records, autopsy, or death certificate have been provided.Based on the limited information provided at this time, no conclusions can be made.The actual date of explant is unknown, as reported by the decedent's attorney "underwent surgery in 2015.Removed part of the mesh", therefore we are using (b)(6) 2015 as a date of explant as well as date of event.Should additional information be provided a supplemental emdr will be submitted.Note: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.Not returned.
|
| |
|
Event or Problem Description
|
|
The following was alleged by the patient's attorney: on (b)(6) 2013: the patient was implanted with a bard/davol phasix mesh for the repair of an incisional hernia.On (b)(6) 2015: the patient started experiencing pain radiating from his abdomen and neuropathy.Thereafter, the patients physicians elected for a repair surgery of the hernia and hernia mesh.Upon entering the abdominal cavity, the patients surgeon noted that the bard/davol phasix mesh had failed and removed part of the mesh.It is alleged that subsequent to the bard/davol phasix mesh implant surgery, the patient suffered substantial pain and suffering and has not been able to received proper care for his esophagus cancer because physicians were unable to place a feeding tube due to the mesh, and subsequently died.
|
| |
|
Additional Manufacturer Narrative
|
|
Currently, it is unknown to what extent the bard/davol phasix mesh device may have caused or contributed to the events as alleged by the patient¿s attorney.The information provided alleges the patient underwent additional "repair surgery of the hernia and the hernia mesh" with removal of part of the mesh.A sample was not provided for evaluation and no lot number has been provided; therefore a review of the manufacturing records is not possible at this time.No medical records, autopsy, or death certificate have been provided.Based on the limited information provided at this time, no conclusions can be made.The actual date of explant is unknown, as reported by the decedent's attorney "underwent surgery in 2015.Removed part of the mesh", therefore we are using (b)(6) 2015 as a date of explant as well as date of event.Addendum: h11: this supplemental emdr is submitted to document additional information provided and to correct the date of event and date of explant.Based on the additional information received, no conclusions can be made.Per medical records review, about 1 year 6 months post implant of phasix mesh, patient was diagnosed with infection, necrotic tissue thereby underwent repair with mesh removal.After 1 year 8 months, the patient passed away on (b)(6) 2017.Per the death certificate, the patient's cause of death was listed as "acute congestive heart failure, atrial fibrillation, coronary artery disease." other significant conditions contributing to death are esophageal cancer.It is unknown to what extent the device may have caused or contributed to the patient death.The instructions-for-use supplied with the device list infection as a possible complication.The warning section of the ifu states "if an infection develops, treat the infection aggressively.Consideration should be given regarding the need to remove the mesh.An unresolved infection may require removal of the device." note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
The following was alleged by the patient's attorney: on (b)(6) 2013: the patient was implanted with a bard/davol phasix mesh for the repair of an incisional hernia.On (b)(6)2015: the patient started experiencing pain radiating from his abdomen and neuropathy.Thereafter, the patients physicians elected for a repair surgery of the hernia and hernia mesh.Upon entering the abdominal cavity, the patients surgeon noted that the bard/davol phasix mesh had failed and removed part of the mesh.It is alleged that subsequent to the bard/davol phasix mesh implant surgery, the patient suffered substantial pain and suffering and has not been able to received proper care for his esophagus cancer because physicians were unable to place a feeding tube due to the mesh, and subsequently died.Addendum per additional information provided: on (b)(6) 2013 - patient was diagnosed with incisional hernia thereby underwent open repair with the implant of phasix mesh.Per operative notes, ¿the hernia sac was identified and dissected.A phasix mesh was placed and secured circumferentially with sutures.¿ on (b)(6) 2015 - patient was diagnosed with infected hernia mesh thereby underwent open repair with the removal of mesh.Per operative notes, ¿sinus was identified and multiple other sinuses within the inferior portion of the incision as well at the umbilicus.The other sinuses were also explored, and each was either noted to contain mesh or prolene suture.Once cleaned, similar findings were noted superiorly to the umbilicus.All necrotic tissue was debrided.The infected mesh and prolene suture had been removed and then the wound was closed with sutures.¿ on (b)(6) 2017 - patient passed away.Per the death certificate, the cause of patient death was listed as ¿acute congestive heart failure, atrial fibrillation, coronary artery disease.¿ other significant conditions contributing to death is esophageal cancer.(not legible another underlying condition).Attorney alleges that the patient had abscess, adhesions, infections, mesh migration, pain and hernia recurrence.
|
| |
|