WILLIAM COOK EUROPE COOK CELECT NAVALIGN FEMORAL VENA CAVA FILTER SET DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number IGTCFS-65-1-FEM-CELECT
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Death (1802); Hematoma (1884); Blood Loss (2597)
Event Type  Injury  
Manufacturer Narrative
(b)(4). Catalog # igtcfs-65-1-fem-celect. (b)(4). This report includes information known at this time. A follow up report will be submitted should additional relevant information become available.
 
Event Description
[pt] allegedly received an implant on (b)(6) 2010 via the right common femoral vein due to pulmonary embolism. [pt] is alleging hematoma, bleeding. It was alleged [pt] expired on (b)(6) 2012 without further details.
 
Manufacturer Narrative
(b)(4). Name and address for importer site: (b)(4). Additional information: investigation is reopened due to additional information provided. It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating "celect - hematoma, bleeding (possible vc perforation), death". Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c. F. R. 803. 56. Unknown if reported patient death is related to filter. Vena cava wall perforation is a known potential complication of vena cava filters. Both symptomatic and asymptomatic events have been reported. Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e. G. , a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter. There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae. E. G. Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae. In contrast, perforation of adjacent organs is reported with clinical sequelae. Unknown if the reported hematoma or bleeding is directly related to the filter. No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e. G. Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.
 
Event Description
No additional information provided at this time.
 
Manufacturer Narrative
(b)(4). Investigation: investigation is reopened due to additional information provided. It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating "celect - hematoma, bleeding (possible vc perforation), death". Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c. F. R. 803. 56. Unknown if reported patient death is related to filter. Vena cava wall perforation is a known potential complication of vena cava filters. Both symptomatic and asymptomatic events have been reported. Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e. G. , a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter. There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae. E. G. Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae. In contrast, perforation of adjacent organs is reported with clinical sequelae. Unknown if the reported hematoma or bleeding is directly related to the filter. No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e. G. Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.
 
Event Description
No additional information provided at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCOOK CELECT NAVALIGN FEMORAL VENA CAVA FILTER SET
Type of DeviceDTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA 4632
Manufacturer (Section G)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
Manufacturer Contact
thomas hessner kirk
sandet 6
bjaeverskov 4632 DK-46-32
DA   DK-4632
56868686
MDR Report Key7586759
MDR Text Key110610514
Report Number3002808486-2018-00650
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090140
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Attorney
Type of Report Initial,Followup,Followup,Followup
Report Date 09/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/11/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date03/02/2013
Device Catalogue NumberIGTCFS-65-1-FEM-CELECT
Device Lot NumberE2541319
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Distributor Facility Aware Date05/22/2018
Event Location No Information
Date Manufacturer Received09/09/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/02/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 06/11/2018 Patient Sequence Number: 1
-
-