| Catalog Number |
7D2648 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No kit lot number was provided for a specific investigation.No sample was returned for further testing, and no additional information was provided.A review of complaint trends revealed that all of the alere determine (b)(6) combo lots are performing according to label claims.The exact root cause of the reported issue could not be determined.The results obtained may possibly be related to the patient sample.The sample may have contained specific substances which may have affected the results.The available evidence suggests that the device is performing within labeled claims.
|
| |
|
Event or Problem Description
|
|
Customer reported (via a customer survey) a false (b)(6) ab result on a serum sample tested with the alere determine hiv 1/2ag/ab combo.Unspecified confirmatory test done at quest was negative.The date of event was not provided.No additional patient information was provided.There is insufficient information to determine if a malfunction occurred.Patient gender, pregnancy status, treatment and patient outcome were unknown.At the time of the event there was no indication of a product malfunction based on the information provided and assessed by alere scarborough, inc.; however, this mdr is a retrospective filing in response to an observation from an fda inspection conducted (b)(6) 2018 at alere (b)(6).
|
| |
|