Fujimoto, yoshihie, et.Al., 'delamination of novel ultrathin bioabsorbable abluminal polymer of platinum chromium everolimus-eluting stent', cardioasc interv and ther 2018 33:97-98, https://doi.Org/10.1007/s12928-016-0442-9.Device is a combination product.(b)(6).(b)(4).Device evaluated by mfr: the complaint device was not received for analysis.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|
It was reported via literature that polymer delamination occurred.The 90% stenosed target lesion was located in the severely calcified mid-left anterior descending coronary artery (lad).After predilatation, delivery of a 24mm synergy stent in the mid-lad was attempted.However, it would not advance to the target lesion in the mid-lad even with the a non-bsc guide extension catheter.Rotational atherectomy was performed.A durable polymer everolimus-eluting stent was then deployed.The final angiogram showed a good result.Scanning electron microscopy demonstrated damage to the bioabsorbable polymer of the synergy stent; most of the damage was delamination of abluminal polymer.It was concluded that delamination of abluminal bioabsorbable polymer may occur more easily and severely when it is delivered through a calcified coronary artery.
|