• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ADVANCE 18 LP LOW PROFILE BALLOON CATHETER; LIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ADVANCE 18 LP LOW PROFILE BALLOON CATHETER; LIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Catalog Number PTA4-18-150-3-20
Medical Device Problem Codes Fluid/Blood Leak (1250); Hole In Material (1293)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Pma=k130293.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported, during a fistulagram procedure, the balloon of the advance 18 lp low profile balloon catheter was found to have a hole in it.This was discovered when the balloon was inflated and contrast was injected.The contrast leaked out of the balloon.The complainant denied any scarring, calcification, or tortuosity of the patient's anatomy.The advance 18 lp low profile balloon catheter was removed and the procedure was successfully completed using another.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Investigation/evaluation: a review of the complaint history, device history record, documentation, drawing, instructions for use (ifu), quality control, specifications, and a visual inspection of the returned device were conducted during the investigation.The visual inspection of the returned device confirmed the presence of a pin hole leak, which was noted at a distance of 20.9 cm from the distal end of the device.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should be noted that there were no other reported complaints for this lot number.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCE 18 LP LOW PROFILE BALLOON CATHETER
Common Device Name
LIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7587878
Report Number1820334-2018-01608
Device Sequence Number13782471
Product Code LIT
UDI-Device Identifier10827002309460
UDI-Public(01)10827002309460(17)210427(10)8846664
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberPTA4-18-150-3-20
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 06/05/2018
Supplement Date Received by Manufacturer10/02/2018
Initial Report FDA Received Date06/11/2018
Supplement Report FDA Received Date10/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age86 YR
-
-