It was reported, during a thrombectomy of a fistula, the wire of the micropuncture transitionless access set "fell apart" where it meets the floppy end.There was no calcification, scarring, tortuous anatomy nor was there angulation of the patient's anatomy reported.The separated device portion was snared and successfully removed from the patient's anatomy.The procedure was completed without any further issue.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
|
Investigation: evaluation: a review of the dimensional verification, complaint history, device history record, documentation, drawing, manufacturing instructions, quality control, specifications, and a visual inspection and functional evaluation of the returned device were conducted during the investigation.The visual inspection of the returned device confirmed the presence of a kink in the wire guide, which occurred a distance of approximately 7 mm from the distal tip.The weld ball was observed to be intact and centered at the distal end.The proximal end of the device appeared to be flattened and not round.Biomatter was present on the device.The mandril was not returned with the rest of the device.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.A failure related subassembly had two nonconformances which were related to the failure mode; however, these units were either reworked or scrapped and not replaced.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be determined.Measures are being conducted to address this failure mode.We will continue to monitor for similar complaints.Per the risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|