| Model Number |
M001712180 |
| Medical Device Problem Code |
Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/18/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Age at time of event: 18 years or older.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that stent profile problem occurred.The 80% stenosed target lesion was located in a shunt in the moderately tortuous, mildly calcified, and 7 mm in diameter brachial vein.An 8x38x75/ iliac 6f wallstent¿ rp endoprosthesis was advanced to treat the lesion.However, it was noted during deployment that the length of the stent was about 1 cm.The procedure was completed with additional wallstent placed.No patient complications were reported and the patient's status was stable.
|
| |
|
Event or Problem Description
|
|
It was reported that stent profile problem occurred.The 80% stenosed target lesion was located in a shunt in the moderately tortuous, mildly calcified, and 7mm in diameter brachial vein.An 8x38x75/ iliac 6f wallstent¿ rp endoprosthesis was advanced to treat the lesion.However, it was noted during deployment that the length of the stent was about 1cm.The procedure was completed with additional wallstent placed.No patient complications were reported and the patient's status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: the stent of the returned device was not returned for analysis as it was implanted in the patient.As per dfu ¿ wallstent¿ rp endoprosthesis, iliac, the size of the stent that is mounted on the returned device before deployment is 38mm.When deployed in a vessel of 7mm the implanted length of the stent is approximately 57mm.A visual and microscopic examination identified no issues with the markerbands.The position of the markerbands were measured as the stent is cut and placed using the markerbands as guidance.The markerbands were noted to be within specification.No issues were noted with the markerbands that could have contributed to the complaint incident.A visual and tactile examination identified no damage or any issue along the length of the device.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is not confirmed - returned as there was no evidence of either the alleged issue(s) or any defect which could have contributed to the event.(b)(4).
|
| |
|