| Model Number |
M001712360 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/19/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age at time of event: 18 years or older.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a foreign matter was found stuck on the device.The 99% stenosed target lesion was located in the mildly tortuous and severely calcified external iliac artery.Following pre-dilation, a 10x49x75/ iliac 6f wallstent¿ rp endoprosthesis was successfully implanted for treatment.However, when the delivery shaft of the device was checked after the procedure was completed, it was noted that a white fluff was sticking on the shaft.No patient complications were reported and the patient's status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr., eval summary attached, method codes, result codes, and conclusion codes updated. device evaluated by mfr.: device was returned for analysis.The wallstent iliac device was returned with the stent fully deployed from the delivery system and a white fibrous material was noted to be attached to the delivery system.A sample of the foreign material was sent for fourier transform infrared radiation (ftir) analysis.Ftir analysis confirmed that the foreign material was a polyethylene terephthalate (polyester) fibre with adhesive attached to it.No issues were identified with the stent holder.The stent cups were noted to be damaged.This type of damage is consistent with excessive force being applied to the device during stent deployment.The shaft of the device was noted to be severely kinked at approximately 235mm distal of the main valve.This type of damage is consistent with excessive force being applied to the device when encountering resistance.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion was unable to be determined. (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a foreign matter was found stuck on the device.The 99% stenosed target lesion was located in the mildly tortuous and severely calcified external iliac artery.Following pre-dilation, a 10x49x75/ iliac 6f wallstent¿ rp endoprosthesis was successfully implanted for treatment.However, when the delivery shaft of the device was checked after the procedure was completed, it was noted that a white fluff was sticking on the shaft.No patient complications were reported and the patient's status was stable.
|
| |
|