| Catalog Number |
368660 |
| Medical Device Problem Code |
Incorrect Or Inadequate Test Results (2456)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported with the use of the bd vacutainer® serum blood collection tubes there was an issue with erroneous results.It was stated ¿the customer reported that the tube looks like it was wet inside that then it dried and left spots all over the tube.¿ there was no report of injury or further medical intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd received samples and photos from the customer facility for investigation.The photos were evaluated and the customer¿s indicated failure mode for additive abnormality with the incident lot was not observed.Additionally, evaluation of the customer samples was performed and additive abnormality was not observed.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Investigation conclusion: based on evaluation of the customer samples, the customer¿s indicated failure mode for additive abnormality with the incident lot was not observed as all samples met the required specifications.Additionally, the photos did not identify a specific product issue.Root cause description: based on the investigation, a root cause could not be determined.The product was found to be in conformance and meet release specifications.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Search Alerts/Recalls
|