| Model Number |
8000.COM03 |
| Medical Device Problem Code |
Device Operational Issue (2914)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/16/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Assuming worst case during surgery, (patient harm), failing of foot pedal during surgery can lead to huge delay in surgery, hence reportable.This reportable event was identified during a voluntary retrospective review of all complaints since 2015 by the manufacturer ((b)(4)).Details of this activity were discussed with cdrh office of compliance ((b)(6) during a tele-conference on (b)(6) 2017.All available information has been disclosed.This complaint has now been closed.
|
| |
|
Event or Problem Description
|
|
Foot pedal stopped working in middle of case while cabled to eva.Customer not able to connect foot pedal wirelessly.Restarted machine, still received error code fop3102 but was able to resolve error code and continue case.New foot pedal will be sent.
|
| |
|